PENGENDALIAN MUTU PEMERIKSAAN KOLESTEROL SERUM KONTROL KOMERSIAL BERDASARKAN WAKTU PENUNDAAN

Authors

  • Fizi Altian Nandasari Universitas ‘Aisyiyah Yogyakarta
  • Titin Aryani Universitas ‘Aisyiyah Yogyakarta
  • Joko Murdiyanto Universitas ‘Aisyiyah Yogyakarta

DOI:

https://doi.org/10.31004/jkt.v5i3.33275

Keywords:

Bahan kontrol, penundaan, kolesterol

Abstract

Each phase in laboratory examination is at risk of contributing errors that can affect the results. Quality control using control materials is useful for ensuring the accuracy of laboratory examination, one of which is cholesterol examination.  This study aims to determine results of quality control with control materials that are immediately examined and delayed 7 hours at room temperature (20-25°C) in cholesterol testing. This study used quantitative methods and quasi experimental design. The sample in this study was commercial control serum and used primary data collection techniques with cross sectional approach. Variables in this study are the length of delay as independent variable and cholesterol levels as dependent variable. Kolmogorov-Smirnov normality test and Independent Sample T-Test significance test were used to analyze the data. Quality control analysis was performed using Levey-Jennings chart and Westgard Rules. The difference between 0 and 7 hours cholesterol examination was 0,44% and the significance test result of p = 0,010 showed that there was significant difference between 0 and 7 hours cholesterol examination. The quality control results of 0-hour cholesterol examination followed Wesgard 12S warning rule and control serum that was delayed for 7 hours at room temperature had results that followed Westgard 12S, 22S and 13S rules. There is a significant difference in cholesterol examination with control serum that is examined immediately and delayed for 7 hours, and delay control serum for 7 hours can cause random and systematic errors.

References

Artha, C., Mustika, A., & Sulistyawati, S. W. (2017). Singawalang Leaf Extract Effects On LDL Levels of Hypercholesterolemic Male Rats. Ejki, 5(2), 105–109.

Badan Litbangkes. (2018). Laporan Riskesdas 2018 Nasional. Lembaga Penerbit Balitbangkes.

Erwinanto, Sunanto, N., Anwar, S., Dwita, R. D., Erika, Renan, S., Sony, H. W., & Yusra, P. (2022). Panduan Tata Laksana Dislipidemia 2022. Perhimpunan Dokter Spesialis Kardiovaskular Indonesia.

Geru, Y. L. (2018). Perbedaan Kadar Kolesterol High Density Lipoprotein (HDL) pada Serum Segera Dan Tunda 4 Jam. Jurnal Universitas Muhammadiyah Semarang, 1, 1–4.

Kemenkes, R. (2013). Peraturan Mentri Kesehatan Republik Indonesia (2013) Nomer 43 Tahun 2013. Penyelenggaraan Laboratorium Klinik Yang Baik. Menteri Kesehatan Republik Indonesia.

Kusmiati, M., Nurpalah, R., & Restaviani, R. (2022). Presisi dan Akurasi Hasil Quality Control Pada Parameter Pemeriksaan Glukosa Darah di Laboratorium Klinik Rumah Sakit X Kota Tasikmalaya. JoIMedLabS, 3(1), 27–37.

Lamik, I. (2018). Perbedaan Kadar Kolesterol Pada Pemeriksaan Serum Segera dan Tunda 4 Jam. Jurnal Universitas Muhammadiyah Semarang, 1, 1–4.

Lestari, A. T. (2019). Aplikasi Pemberian Susu Kedelai dan Jahe Terhadap Penurunan Kadar Kolesterol Pada Penderita Penyakit Jantung Koroner. Karya Tulis Ilmiah. Universitas Muhammadiyah Magelang.

Lestari, W. S., Karwiti, W., Latifah, A., Listiani, Y., & Harianja, S. H. (2022). Sera Pooled Stability As a SGPT Control Material With Storage Time and Temperature Variation. Journal of Medical Laboratory and Science, 2(1), 33–39.

Mufaridah, L., & Aryani, T. (2022). Analisis Kadar Kolesterol dan Trigliserida pada Serum Kontrol Komersial Berdasarkan Lama Penyimpanan. Skripsi. Universitas 'Aisyiyah Yogyakarta.

Noviyanti, T., & Ardiningsih, P. (2013). Pengaruh Temperatur Terhadap Aktivitas Enzim Protease Dari Daun Sansakng (Pycnarrhena cauliflora Diels). Jurnal Kimia Khatulistiwa, 1(1), 1–6.

Purbayanti, D. (2015). Pengaruh Waktu Pada Penyimpanan Serum Untuk Pemeriksaan Kolesterol Total. Jurnal Surya Medika, 1(1), 1–17.

Ramdan, A., & Aryani, T. (2023). Kontrol Kualitas Serum Kontrol Komersial Pemeriksaan Glukosa Berdasarkan Lama Penyimpanan. Jurnal Kolaboratif Sains, 6(11), 1384–1391.

Salma, F. D., Rahayu, I. G., Kurnaeni, N., & Rinaldi, S. F. (2017). Cost-Effectiveness Analysis (Cea) Bahan Kontrol Komersial Dan Pool Serum Pasien. Jurnal Riset Kesehatan, 11(1),

Siregar, M. T., Wulan, W. S., Setiawan, D., & Nuryati, A. (2018). Bahan Ajar Teknologi Laboratorium Medik (TLM): Kendali Mutu. Badan Pengembangan dan Pemberdayaan Sumber Daya Manusia Kesehatan.

Yaqin, M. A., & Arista, D. (2015). Analisis Tahap Pemeriksaan Pra Analitik Sebagai Upaya Peningkatan Mutu Hasil Laboratorium di RS. Muji Rahayu Surabaya. Jurnal Sains, 5(10), 1–7.

Downloads

Published